Items list
Description
The high-performance ADVIA Centaur® XP system has more onboard reagents and dedicated STAT capabilities to maximize productivity, regardless of volume or types of tests. Always ready, continuous operation without interruption.
- High throughput, up to 240 tests/hour
- Intuitive software for ease of operation
- STAT port guarantees priority sampling at any time
- Data archive reduces administrative tasks
- Extensive menu with 30 onboard reagents
- Increase productivity by connecting to Siemens' Aptio™ Automation, ADVIA® Automation, StreamLAB® Automation or VersaCell® Solutions
Comprehensive Menu
- Comprehensive diagnostic testing covers screening, diagnosis, risk assessment and monitoring
- 30 onboard reagents maximize test capacity
Powerful Productivity
- High throughput, up to 240 tests per hour, to keep pace with peak workload times
- High-resolution touch screen for easy interaction
- Expanded ancillary reagent and fluid capacity
- Data archive feature reduces time for administrative tasks
- Universal sample rack design eliminates manual tasks
Excellent Patient Care
- STAT port guarantees priority sampling at any time
- Disposable pipette tips eliminate sample-to-sample carryover
- Automatic dilution, repeat and reflex testing
- Smart Algorithm Software automatically repeats and confirms reactive testing
Superior Service
- 360º status light displays system information visible from all points
- User interface adjusts for customized ergonomics
- No daily startup with automated maintenance features
- New instrument sensors enhance remote diagnostic functions
Overview | |
System Description | Random access immunoassay system |
Throughput | Up to 240 tests per hour |
Time To First Result | 18 minutes; results every 15 seconds thereafter, assay dependent |
Assays On Board | 30 |
Assay Packs | ADVIA Centaur® ReadyPack™ assays |
Assay Technology | Direct Chemiluminescence using Acridinium Ester technology |
Sample Handling | |
Sample Tubes | 1 mL, 2 mL, 3 mL, 5 mL, 7 mL, 10 mL, micro container tubes |
Sample Carriers | Universal 5-position rack holds multiple tube types; 180 samples fully loaded; no-pause loading and unloading |
STAT Handling | Dedicated STAT port accepts samples any time |
Sample Types | Serum, plasma, urine - assay dependent |
Sample Volume | 10 uL to 200 uL assay dependent |
Sample Integrity Control | Pressure transducer technology with clot management, detection, and flagging. |
Sample Pipette Tips | Ensure zero sample carryover |
Automatic Dilutions | Varies by assay, up to 1:1000 |
Bar Codes | |
Code 128, Code 39, Codabar, Interleaved 2 of 5 | |
Reagent Handling | |
Reagent Cooling | 30-position reagent tray cooled 4°C to 8°C |
Reagent Integrity Control | Barcode reagent identification, automatic inventory tracking and flagging, calibration validity tracked and flagged, reagent on board stability tracked and flagged, reagent expired/ reagent low flagging |
Reagent Mixing | ReadyPacks automatically rocked onboard |
Ancillary Reagents | 25-position ancillary reagent tray cooled 4°C to 8°C |
Supplies | |
Cuvettes | 1,000 ADVIA Centaur cuvette capacity |
Sample Tips | ADVIA Centaur type, 840 on board, automatic tracking, and flagging of inventory |
Continuous Operation | |
Loading/unloading of samples and consumables at anytime without he need to interrupt system operations, includes: sample racks, STATS, reagent packs, ancillary reagent packs, reaction cuvettes, sample pipette tips, calibrators, controls, wash fluids, water, and waste. | |
User Interface | |
Monitor | 19-inch diagonal high-resolution LCD touch screen with adjustable height |
Operating System Results Management | Dual Sun Microsystems Sparc® II processors; Microsoft® Windows® XP |
Host Interface | RS232C bidirectional and host query; Ethernet connection available; 25,000 result storage; archive to removable media; automatic reflex testing |
Smart Algorithm Software | Repeats and confirms reactive testing; results available as completed |
Quality Control | |
Advanced QC package includes Levy-Jennings plots and Westgard rules; 65,000 control results can be stored | |
Removable Media | |
DVD or CD-RW | |
Device Management | |
100 Base-T Ethernet port for remote access and Siemens Remote Service proactive service modem | |
General Specifications | |
Electrical | 208-240 V at 50/60 Hz; meets regulatory requirements for UL, CAN/CSA, IEC1010. JIS, CE, CISPR11 Class A |
Power Consumption | 1 KW in ready state; 1.5 KW in operating state |
Water | Deionized type II |
Dimensions | 51.5 (h) x 72.4 (w) x 41.0 (d) in-130.9 (h) x 183.9 (w) x 104.2 (d) cm |
Weight | 1,200 lbs (545 kg) |
Floor Load-bearing Requirement | 390 kg/m2 (80lb/ft2) |
Operating Environment Temperature | 18-30OC; humidity 20-80% noncondensing |
Ventilation | 10,000 BTU/hour |
Noise Emission | 58.8 db in ready state; 61.3 in operating state |